Wincom Company Ltd.
Main products:Microscope,centrifuge,water bath,distiller,stirrer hotplate,photoelectric colorimeter,VIS spectrophotometer,pH meter,balance,scale,educational meter,electrophoresis,incubator,oven,laboratory glassware and porcelain,operating lamp,gas chromatograph,educati
Products
Contact Us
  • Contact Person : Ms. Huang Judy
  • Company Name : Wincom Company Ltd.
  • Tel : 86-731-88660686
  • Fax : 86-731-84121122
  • Address : Hunan,CHANGSHA,NO.46,MINGCHENG INTERNATIONAL,RENMING ROAD(E)
  • Country/Region : China
  • Zip : 410005

High Accuracy PH Meter

High Accuracy PH Meter
Product Detailed
Specifications: Range : 0.00 to 14.00 pH Resolution : 0.01 pH Calibration: manual, 2 point Environment : RH 95% max

Technical Parameters

Model

PH-98107

PH-98108

Range

0.00~14.00PH

Resolution

0.01 pH

Accuracy

±0.10pH

±0.05pH

Calibration 

2 point

A T C

0 to 60°C

Power

4×1.5v alkaline

Dimensions

177×42×24mm

Weight

90g

Calibration :

Immerse the tester up to the minimum immersion level in pH 6.86(or 7.01) buffer solution.

Allow the reading to stabilize with the supplied screwdriver, adjust the PH7 calibration trimmer to

6.86(or7.01)pH..

Rinse the electrode with top water immerse it into the second buffer.

Use pH4.00 if the solution to measured is acidic(pH<7),or pH9.18( 10.0) for alkaline samples(pH>7).

Allow the reading to stabilize and with the screwdriver adjust the PH4/10 calibration trimmer to read 4.00 or 9.18(or10.01) pH.

Operation : 

Remove the protective cap. Do not be alarmed if white crystals appear around the cap. This is normal with pH electrodes and they dissolve when rinsed with water.

Turn the meter on sliding the switch on the top.

Immerse into solution up to the max immersion level.

Stir gently and wait until the display stabilizes.

After use, switch off pH meter, use the water clean the electrode and replace the protective cap.

DO NOT USE DISTILLED OR DEIONIZED WATER FOR STORAGE PURPOSES.

Large fluctuations in readings could be due to lack of calibration, dry electrode or rundown batteries.

High Accuracy PH Meter



Copyright Notice @ 2008-2022 ECVERY Limited and/or its subsidiaries and licensors. All rights reserved.